D500

Mediana

Biphasic defibrillation

  Biphasic defibrillation , pacing and complete monitoring in on portable device

- Multifunctional Defibrillator/monitor

- manual and AED Operation

- NON-invasive pacing mode

- Advanced Biphasic Technology

- Defibrillation with paddles

- 12 lead ECG monitoring 

screen size:170.0*128mm(8.4 in diagonally across the TFT-LCD screen)

screen type/color: liquid crystal display (LCD)color

resolution:800*600 pixel

CONTROLS

standard knob mode key (off,AED,manual ,pacing and monitor ):11 buttons(shock,select energe leavel charge ,analyze,NIBP,LEAD,alarm,size,print,rate.ma):5 soft key

ALARMS:

categories:patient status and system status

priorities:low,medium and high priorities

notification:audible and visual

setting:default and individual

alarm volume level:45 to 85db

physical characteristics and printer

instrument

dimensions :340*305*210mm(W*H*D) including a battery excluding paddles,options and accessories

weight:6.16 kg including battery excluding paddles options and accessories

ECG:type CF with defibrillation protection

SPO2:type  CF with defibrillation protection

Temperature::type  CF with defibrillation protection

Etco2::type  CF with defibrillation protection

NIBP:type  CF with defibrillation protection

IBP::type  CF with defibrillation protection

PADDLE::type  CF with defibrillation protection

Mode of operation:continous

printer

type;thermal

weight190g

Number of channels:1 to 3  channels

paper width 80mm

printer speed :25mm/s

ELECTRICAL

instrument

power requirement::AC mains100 to 240v,50/60hz  , 60to160va DC mains 18vdc,7.0a with DC/DC adapter

model:MDD150-1218(MDD150-1218;INPUT:12-16VDC,160VA,QUTPUT::18vdc,7.0A)

battery (option)

type:Li-ionbattery

voltage 14.4v/6600mah

discharge :A minimum of 200 shocks at 200 joules(per battery)

operating time:5 hours (per battery) AT the following condition:no printing ,no external communication,no audible alarm sound and room temperature:25°c 

recharge:5 hours with D500 turned on/off

  Duai Battery:automatic switching

Defibrillator

biphasic waveform :biphasi truncated exponential resuscitation guidelines:selectable AHA/ERC

MANUAL MODE

shock energy leavel:external paddles

1,2,3,4,5,6,7,8,9,10.15,20,30,40,50,75,100,125,150,175,200,300,360

automatic discharge time :60 seconds

charging time to 200j:within 6 seconds with rated main voltage/dcmain voltage (battery within7seconds

charging time to 360j: within8seconds with rated main voltage/dc main voltage(battery within9seconds

synchronous cardioversion:energy transfer begins within 60 msec of the QRS peak

AED mode

1ch ECG measurement

lead   lead II

patient impedavnce 25 to 175 ohm

heart rate:20 to 300 bpm

charging time to200  : within6 seconds with rated main voltage/DC main voltage(battery within7 seconds

Delivered energy

The D500 delivers shocks to load impedances from25 to 175 ohm

the duration of each pulse of the waveform is dynamically adjusted based on delivered charge in order to compensate for patient impedance variation as shown below

load resistance (ohm) Delivered energy

25       203

50       198

75       200

100     199

125     198

150     197

175     197